EasyManua.ls Logo

A&D FX-2000i - Details of the Function Table

A&D FX-2000i
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
10.3. Details of the Function Table
Class
Item and Parameter
Description
ba5fnc
Environment
Display
Cond
Condition
0
Fast response, sensitive value FAST
With “Hold 1, sets the
averaging time.
1
MID.
2
Slow response, stable value SLOW
5t-b
Stability band width
0
Stable range is ±1 digit
The stabilization indicator
illuminates when the display
fluctuation is within the range.
With Hold 1, sets the
stabilization range.
1
2
Stable range is ±3 digits
Hold
Hold function
0
OFF
Holds the display when stable in
animal mode. With Hold 1,
ANIMAL turns on.
1
ON
trc
Zero tracking
0
OFF
Keeps zero display by tracking
zero drift.
1
Normal
2
Strong
3
Very strong
5pd
Display refresh rate
0
5 times/second
Period to refresh the display
1
10 times/second
2
20 times/second
pnt
Decimal point
0
Point (.)
Decimal point format
1
Comma (,)
p-on
Auto display-ON
0
OFF
Turns on the weighing mode
display when the AC adapter is
connected.
1
ON
poff
Auto display-OFF
0
OFF
Turns off the display after 10
minutes of inactivity.
1
ON (10 minutes)
rng
Display at start
0
Displays
Select whether or not to display
the minimum weighing value at
weighing start.
1
Does not display
beep
Beep
0
Does not sound
Select whether or not to sound
the beep when operating on
keys.
1
Sounds
Cl adj *
Clock
Refer to 10-9 Clock and calendar
function”.
Confirms and sets the time and
date. The time and date are
added to output data.
Cp fnc
Comparator
Cp
Comparator mode
0
No comparison
1
Comparison, excluding "near zero" when stable value or
overloaded
2
Comparison, including "near zero" when stable value or
overloaded
3
Continuous comparison, excluding "near zero"
4
Continuous comparison, including "near zero"
bep
-
LO buzzer
0
OFF
1
ON
bep-
OK buzzer
0
OFF
1
ON
bep
-
HI buzzer
0
OFF
1
ON
:
Factory settings. Digit is a unit of minimum weighing value
*
:
Only for the FZ-i
/
FZ-iWP
/
FZ-GD series

Table of Contents

Related product manuals