8.5. Correcting the Internal Mass Value: 1 (Only for FZ-
i
/
FZ-
i
WP
/
FZ-GD series)
The FZ-i / FZ-iWP / FZ-GD series balance can correct the internal mass value using an external weight.
Method 1: Calibrate the balance using the internal mass. Measure the external weight. Calculate
the correction value. And store it in the balance.
The adjustable range of the internal mass value is as follows:
Correction reference value
FZ-120i FZ-120iWP FZ-120GD
-0.050 g to
+0.050 g
FZ-200i FZ-200iWP FZ-200GD
FZ-300i FZ-300iWP FZ-300GD
200 g
FZ-1200i FZ-1200iWP FZ-1200GD
-0.50 g to +0.50 g
FZ-2000i FZ-2000iWP FZ-2000GD
FZ-3000i FZ-3000iWP FZ-3000GD
2 kg
Operation
Example : Correcting the weight value by +0.003 g in 100 g using the FZ-120i
If correcting the weight value by +0.003 g in 50 g, use the correction value of +0.006 g as
the correction reference value is 100 g..
1 Perform calibration using the internal mass (one-touch
calibration). Place the external weight on the FZ-120i weighing
pan to obtain the correction value.
2 With the factory setting, the balance can not correct the internal
mass value. Refer to “9.
FUNCTION SWITCH AND
INITIALIZATION” and set the internal mass value correction switch to
.
3 In the weighing mode, press and hold the SAMPLE key to
display ba5fnc ..
4 Press the SAMPLE key several times until C5 in 1 is displayed.
If C5 in 1 is not displayed, perform step 2.
5 Press the PRINT key.
Correct the internal mass value using the following keys:
RE-ZERO key To increase the value by one.
MODE key To decrease the value by one.
PRINT key To store the correction value
and display the next menu item
of the function table.
CAL key To cancel the operation and return to the
weighing mode.
6 Press the CAL key to return to the weighing mode.
7 Press the CAL key to perform calibration using the internal mass.
8 Confirm that the internal mass value has been corrected properly
by placing the external weight on the weighing pan.
If the internal mass value has not been corrected properly,
repeat the procedure to adjust the correction value.
Correct the internal mass by
+0.003 g in 100 g. Calibrate with
the corrected internal mass.
50 digits after +50 digit
Press and
ho
ld
th
e
k
ey
(for 2 seconds)