EasyManua.ls Logo

A&D GF-30K - Page 94

A&D GF-30K
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92
GF-8K GF-8K2 GF-10K GF-12K GF-20K GF-30K GF-32K
Weighing capacity
8.1 kg 10.1 kg 12 kg 21 kg 31 kg 31 kg
8.1008 kg 31.008 kg
Maximum display
8.10084 kg
2.10009 kg *1
10.10084 kg 12.0084 kg 21.0084 kg 31.0084 kg
6.1009 kg *1
Minimum weighing value
(1 digit)
0.01 g 0.1 g / 0.01 g 0.01 g 0.1 g 1 g / 0.1 g
Repeatability
(Standard deviation)
0.01 g 0.05 g / 0.01 g 0.01 g 0.1 g 0.5 g / 0.1 g
Linearity
±0.03 g ±0.1 g/ ±0.02g ±0.03 g ±0.2 g ±1 g / ±0.2 g
Stabilization time
(Typical at
FAST
)
Approx. 1.5 seconds
Sensitivity drift,
(10°C ~ 30°C / 50
°
F ~ 86
°
F)
±2 ppm/°C ±3 ppm/°C ±2 ppm/°C ±3 ppm/°C ±5 ppm/°C
Operating environment
5°C to 40°C (41°F to 104°F), 85%RH or less (No condensation)
Internal mass
not available
Weighing data of data
memory
40 data
Time and clock function
not available
Display refresh rate
5 times/second or 10 times/second
Minimum unit
mass
0.01 g 0.1 g 0.01 g 0.1 g 1g
Counting
mode
Number of
samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
1 g 10 g 1 g 10 g 100g
Percent
mode
Minimum 100%
display
0.01 %, 0.1 %, 1 % (Depends on the reference mass stored.)
Interface
(Provided as standard)
RS-232C with Windows Communication Tools Softare WinCT
External calibration weight
2 kg, 3 kg, 4 kg, 5 kg,
6 kg, 7 kg, 8 kg
2 kg, 3 kg,
4 kg, 5 kg,
6 kg, 7 kg,
8 kg, 9 kg,
10 kg
5 kg, 10 kg 10 kg, 20 kg 20 kg, 30 kg
Weighing pan
270 x 210mm
External dimensions
300(W) x 355(D) x 111(H) mm
Power supply &
AC adapter type
Power consumption: Approx. 11VA (supplied to the AC adapter )
Confirm that the adapter type is correct for the local voltage and power receptacle type.
Weight
Approx. 8.3kg
Dust and water protection
Complying with IP65
*1: The balance allows weighing using a precision range, even with a heavy tare placed on the pan. (Smart range function)

Table of Contents