EasyManua.ls Logo

A&D GF-30K - Details of the Function Table

A&D GF-30K
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
10-2. Details Of The Function Table
Class Item and Parameter Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by response
adjustment. With "
Hold 1
", sets the
averaging time.
0
Stable when within
±
1 digit
1
5t-b
Stability band width
2
Stable when within
±
3 digits
The stabilization indicator illuminates
with the display fluctuation within the
range. With "
Hold 1
", sets the stable
range.
0
OFF
Hold
Hold function
1
ON
Holds the display when stable in
animal mode. With "
Hold 1
",
ANIMAL
turns on.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
5pd
Display refresh rate
1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
p-on
Auto display-ON
1
ON
Turns on the weighing mode display
when the AC adapter is connected.
0
OFF
poff
Auto display-OFF
1
ON (10 minutes)
Turns off the display after 10
minutes of inactivity.
0
OFF
g5i
Capacity indicator
1
ON
Capacity indicator.
Zero: 0%
Maximum capacity: 100%
0
OFF
add
Accumulation function
1
ON
Displays and outputs the total value
of the weighing data.
0
Does not display
ba5fnc
Environment
Display
rng
Display at start
1
Displays
Select whether or not to display the
smallest displayable weighing value
at weighing start.
Cl adj
Clock
Refer to "10-7. Clock and Calendar Function"
Confirms and sets the time and
date. The time and date are added
to the output data.
0
No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded
3
Continuous comparison, excluding "near zero"
Cp
Comparator mode
4
Continuous comparison, including "near zero"
0
Set the upper lower limit value
Select
Cp Hi
or
Cp lo
.
Cp in
Data input method
1
Set the reference value
Select
Cp ref
or
Cp lmt
.
0
Not added
Cp-r
Comparison results
1
Added
Select whether or not to add the
comparison results to the output
data.
0
OFF
Cp-b
Main display comparison
1
ON
Displays the results on the main
portion of the display in place of the
weight value.
0
OFF
bep
LO buzzer
1
ON
Select whether or not to sound the
LO buzzer.
0
OFF
bep
OK buzzer
1
ON
Select whether or not to sound the
OK buzzer.
0
OFF
Cp fnc
Comparator
bep
HI buzzer
1
ON
Select whether or not to sound the
HI buzzer.
: Functions for GX-K series.
:
Factory settings. Digit is a unit of minimum weighing value.
Displayed
only when
Comparator
output
(GX-04K) is
installed

Table of Contents