19
5.3.
Principal Switch Operations
5.3.1.
A-BPM Operations
To start or suspend A-BPM.
To switch between A-BPM and S-BPM
Step
1.
Store the preset program (of start times and intervals) for A-BPM.
Step
2.
Set the AUTO switch for the following operations.
"ON"
.......
A-BPM is started and the mark is shown.
Blood pressure measurements are performed in
accordance with preset A-BPM program.
"OFF"
......
A-BPM is suspended and the
mark is hidden.
Blood pressure measurement can be performed using
the preset S-BPM programs.
To Lock A-BPM to "ON".
Keep the AUTO switch to "ON" using the LOCK switch so
that A-BPM can be performed.
To expand A-BPM interval time.
Step
1.
Set the sleep mode to "ON" before the measurement.
Step
2.
Set the AUTO switch to "ON" to use A-BPM.
The mark is shown.
Step
3.
When the EVENT switch is pressed during A-BPM, the interval
time is doubled.
When the EVENT switch is pressed again, the interval time
returns to basic value.