30
6.2.
Self Blood Pressure Measurement (S-BPM)
Set the AUTO switch changes to "OFF" to use S-BPM programs.
The recorder is equipped with the S-BPM five types of programs in
accordance with varying measurement environment conditions.
Parameters and measurement results can be stored in the memory.
Name Descriptions & actions of the program Items
OBP
Office blood pressure
Program for blood pressure measurement by hospital
staff. "1 scope" : Blood pressure measurement is
occurs once.
N.A.
AOBP
Automated office blood pressure
Program for blood pressure measurement after resting
inside the hospital. "1 scope" : Measurement is
performed using measurement count and interval time.
Count
Interval
HBP
Home blood pressure
#1
Program for blood pressure measurement at home.
"1 scope" : Measurement is performed using
measurement count and interval time.
Count
Interval
ANBP
Automated night blood pressure
#2
Program for blood pressure measurement during night.
It uses measurement count and interval time. The
ANBP can specify up to six start times a day.
Start time
Count
Interval
ASBP
Automated self blood pressure
#2
Program that indicates start times with sound from the
buzzer. Press the
/ switch to measure the
blood pressure at home. The buzzer can specify up to
six times a day.
Start time
of alarm
Count
Interval
#1
: Blood pressure is measured in accordance with the Japanese
Society of Hypertension.
#2
: When the measurement count and interval time of ANBP or
ASBP are changed, HBP settings are also changed.