EasyManua.ls Logo

A&D TM-2441 - Initial Pressurization Value; A-BPM Preset Programs

A&D TM-2441
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
8.2.3.
Initial Pressurization Value
An initial pressurization value can be set beforehand.
160, 180, 210, 240, 270, AUTO [mmHg]
If AUTO is specified as the initial pressurization value, a proper
pressurization value is selected automatically. Factory setting is
180 mmHg.
Step 1.
Set the AUTO switch to "OFF".
Step 2.
If the indication of the display is hidden, press the START/STOP
or EVENT switch to return to the display of waiting mode.
Step 3.
Select a pressurization value using the START/STOP switch.
8.3.
A-BPM Preset Programs
The initial settings can be configured using the following methods.
The method to use switches on the recorder.
The method to use
dedicated peripheral
that is connected to the
recorder using the USB cable.
A-BPM can use only while the automated measurement can be
performed.
AUTO switch to "OFF
"
S-BPM program
Pressurization value
Current time

Table of Contents

Other manuals for A&D TM-2441

Related product manuals