8.7.2.
S-BPM Operations
Note
Set the built-in clock and initial pressurization value before
measurement because they are used for S-BPM. Refer to
"8.2.2. The Clock and the Monitor Function of
Measurement" and "8.4. S-BPM Programs".
Remove batteries when the patient removes the recorder and
cuff when ANBP or ASBP is used (even during the waiting
mode). If batteries are left in the battery compartment, the cuff
may break when the recorder inflates the cuff at the next "1
scope". If the patient resumes measurement, insert batteries
and press the
/ switch.
Manual blood pressure measurement can be performed during
S-BPM waiting mode.
The measurement result of the manual blood pressure
measurement can be stored in the memory.
When S-BPM is stopped, the error code E07 is displayed on
the OLED display and stored in the memory.
To start S-BPM
Step 1.
Set the AUTO switch to "OFF".
Step 2.
Operations are as follows:
S-BPM Programs Operations
Office blood pressure OBP
Automated office blood pressure AOBP
Home blood pressure HBP
Press the
/ switch to
start the preset program
during waiting mode.
Automated night blood pressure ANBP
Automated self blood pressure ASBP
Preset program enters
standby until the "start time"
or "start time of alarm".