70
To measure blood pressure during S-BPM immediately.
(
Manual blood pressure measurement
of S-BPM )
Step 1.
If the indication of the LCD panel is hidden, press the
START/STOP or EVENT switch to return to the display of
S-BPM waiting mode.
Step 2.
Press the / switch during S-BPM waiting mode.
The blood pressure measurements of "1 scope" are performed
immediately.
To stop or Suspend S-BPM
Operations are as follows:
S-BPM Programs Operations
Office blood pressure
OBP
Automated office
blood pressure AOBP
Home blood pressure
HBP
Press the
/ switch to stop blood
pressure measurement.
Automated night
blood pressure ANBP
Automated self
blood pressure ASBP
Press the
/ switch to stop blood
pressure measurement. At next start time, blood
pressure is measured or the buzzer sounds.
#1
If you need to stop the recorder completely,
remove batteries from the recorder or switch to
O B P, A O B P, o r H B P.
#1
: Refer to "6.2.1. S-BPM Programs".