35
T : Interval time
6.2.2.
S
-
BPM Waiting Mode
Operation mode and pressure value are displayed on the LCD
panel during S-BPM waiting mode.
#
The indication of the LCD panel will be hidden automatically after
no-operation is continued for few minutes. Even if the indication
of the LCD panel is hidden, the Manual blood pressure
measurement can be started when the
/ switch is
pressed. When the START/STOP or EVENT switch is
pressed, the indication is resumed.
"1 scope" consists of a set of blood pressure measurements
and interval times which are repeated to the number of the
measurement count. The last interval time is omitted.
Clock mark
blinks on the LCD panel during an interval time of
"1 scope".
S-BPM waiting mode is a state that blood pressure is not
being measured during the "interval time" and between the last
"1 scope" and next "1
scope".
Program of the S-BPM
Pressurization value
Current time
Press the / switch to start "1 scope".
T
Time
T T
T
Not measured
Press the / switch to stop "1 scope".
1 scope 1 scope
Start time of alarm