67
8.7.
Blood Pressure Measurement Operations
8.7.1.
A-BPM Operations
When A-BPM is started, blood pressure is measured in
accordance with the preset parameters.
Note
Set the built-in clock and initial pressurization value before
measurement because A-BPM uses them. Refer to "8.2.2. The
Clock and the Monitor Function of Measurement" and "8.3.
A-BPM Preset Programs".
When the recorder is removed, set the AUTO switch to "OFF".
If the recorder is removed during A-BPM, the inflation of the cuff
is started next start time, the cuff may break. When A-BPM is
resumed, set the AUTO switch to "ON".
The mark is displayed while A-BPM is used.
Manual blood pressure measurement can be performed during
A-BPM waiting mode.
The measurement result of the manual blood pressure
measurement can be stored in the memory.
When A-BPM is stopped, the error code E07 is displayed on
the OLED display and stored in the memory.
To start A-BPM
Step 1.
Set the AUTO switch to "ON".
Step 2.
The mark is shown on the OLED display and LCD panel.
A-BPM is started.