50
Procedure of operation using switches
Step 1.
Set the AUTO switch to "OFF".
Step
2.
If the indication of the display is hidden, press the
START/STOP or EVENT switch to return to the display of
waiting mode.
Step 3.
While holding START/STOP switch, press and hold the
EVENT switch for 3 seconds or more. Sleep will display on
the OLED display.
Step 4.
Specify the sleep mode using the following switches.
If sleep mode is "ON", proceed to step 5.
EVENT switch
.................
Change of the current parameter.
START/STOP switch
......
Decision, next item.
Step 5.
Specify the start time and interval up to six sections using the
following switches.
EVENT switch
.................
Change of the current parameter.
START/STOP switch
......
Decision, next item.
Step 6.
Specify the start time and operation time of the automated
measurement using the following switches.
EVENT switch
.................
Change of the current parameter.
START/STOP switch
......
Decision, next item, end of the settings.
Step 7.
After completing settings, the recorder returns to waiting mode.
Do not remove batteries while charging the settings.
If batteries are removed, input settings again.

Table of Contents