To set the program for S-BPM.
Step
1.
Set the AUTO switch to "OFF".
Step
2.
While holding the / switch, hold the START/STOP
switch for 3 seconds or more. is displayed on the LCD.
Step
3.
Operations are as follows:
Refer to "8.4. S-BPM Programs".
/ switch
....................
Change the current parameter.
START/STOP switch
......
Decision, next item, end of settings.
To measure the blood pressure during S-BPM immediately.
(
Manual blood pressure measurement
of S-BPM )
Step 1.
If the indication of the LCD panel is hidden, press the
START/STOP or EVENT switch to return to the display of
S-BPM waiting mode.
Step 2.
Press the / switch during S-BPM waiting mode.
The blood pressure measurements of "1 scope" are performed
immediately.
"1 scope" consists of a set of blood pressure measurements and
interval times which are repeated to the number of the
measurement count. The last interval time is omitted.
S-BPM waiting mode is a state that blood pressure is not
measured during the interval time and between the last "1
scope" and next "1 scope".