27
The
mark is displayed on the LCD panel while A-BPM is used.
Blood pressure is measured automatically at the A-BPM start time.
An initial pressurization value can be set beforehand.
160, 180, 210, 240, 270, AUTO [mmHg]
If AUTO is specified to the initial pressurization value, the
pressurization value is selected automatically. Factory setting is
180 mmHg. Refer to "8.2.3. Initial Pressurization Value" for
information on how to select an initial pressurization value.
If the first pressurization is not enough, re-pressurizations are
performed automatically up to two times.
When you delete data in the memory or move the AUTO switch
to "OFF", the pressurization value is reset to the initial
pressurization value.
When a measurement error occurs and the waiting time until the
next start time is longer than 8 minutes, blood pressure is
measured once after 120 seconds. The measurement result is
stored in the memory.
If you want to suspend A-BPM, release the LOCK switch and
move the AUTO switch to "OFF".