52
#1
: When the sleep mode is set to "ON", A-BPM is used the Start time
and Operation time of the automated measurement. The Interval
time of these sections (1 to 6) cannot use.
#2
: When sleep mode is set to "OFF", the Interval time is not displayed.
#3
: Example for automated measurement.
Start time : Stores a time. (0 to 23
hour
)
Operation time : Set to "OFF"
Response : A-BPM starts the blood pressure measurement
at the preset Start time and continues until the
AUTO switch is set to "OFF".
#4
: Example for automated measurement.
Start time : Set to "OFF"
Operation time : Stores time to be continued. (1 to 27
hours
)
Response : A-BPM starts blood pressure measurement
and stops after the Operation time.
The content of the item
Sleep mode :
The Interval time for the automated measurement can be specified.
The Interval time of section 1 to 6 cannot use. Refer to "
6.1.2. Sleep
Function and Interval time
".
Section :
24 hours can be separated to six sections in maximum. Each section
can specify the Start time and Interval. A-BPM can use only while
the automated measurement can be performed.
Automated measurement :
The whole of A-BPM can be controlled. Specify the Start time and
Operation time. Refer to "
8.3.2. A-BPM Program Examples
".