78
2600T SERIES |  PRESSURE TRANSMITTERS |  OI/266/FF-EN REV. E
Cat. Error Block err HMI code PRTB_x_VALUE Status
Process errors PRIMARY SECONDARY TERTIARY QUATERNARY PDP_TRIM SP_TRIM
Primary 
Variable Out 
of Range
High 
Range
PRTB.Other  GOOD_NC Non Specific No Limit
UNCERTAIN 
engineering unit 
range violation 
High Limit
GOOD_NC Non Specific No 
Limit
Low 
Range
UNCERTAIN 
engineering unit 
range violation 
Low Limit
PILD Output
PILD 
Affect 
ADTB.Device 
Needs 
Maintenance Soon
GOOD_NC Non Specific No Limit
PILD 
Affect 
BAD Sensor 
Failure No 
Limit
GOOD_NC Non 
Specific No 
Limit
BAD Sensor Failure No Limit
PILD 
Changed 
Operating 
Conditions
PILD 
Affect 
ADTB.Device 
Needs 
Maintenance Soon
GOOD_NC Non Specific No Limit
PILD 
Affect 
BAD Sensor 
Failure No 
Limit
GOOD_NC Non 
Specific No 
Limit
BAD Sensor Failure No Limit
Out Of Service
PRTB.Out of 
Service
BAD Out of Service
Cat Error HMI code Description Possible Cause Suggested Actions
Electronics errors
Memory Failure 
The device data loaded at the start 
up are corrupted precluding the 
correct functionality of the device
Electronic memory corrupted The electronics must be replaced
Electronic Interface 
error.
Exchange of data between 
Electronics and Sensor have 
problems
Exchange of non-critical data 
between sensor and electronics is 
precluded due to problem in the 
transmitter circuit of the electronics 
or in receiver circuit of the sensor
Power cycle the device and retry the 
operation, if the error persist the 
electronics should be replaced
Non-Volatile memory 
burn error
The device continue to work without 
problems but at the next power 
cycle the new configuration will be 
lost
Writings to the electronic non-
Volatile Memory was not successful
The electronics should be replaced 
as soon as possible.
Pressure sensor errors
Sensor 
Invalid
Missing 
Primary 
Signal
The primary signal of the sensor is 
no longer available. The transducer 
is not in a condition to generate a 
valid signal.
The sensor signal is not being 
updated correctly as a result of an 
electronics failure, sensor error or a 
poorly connected sensor cable.
Check cable connection, check 
sensor and if problem persists, the 
sensor must be replaced.
Invalid 
Sensor
Sensor and/or the connected 
electronics are incompatible
The sensor model/version is not 
longer compatible with the 
connected electronic version
The sensor must be replaced
Sensor Memory Fail 
The data in the sensor memory are 
corrupted precluding the correct 
functionality of the device
Sensor memory corrupted The Sensor must be replaced
P-dP Sensor Fail 
The sensor signal value is incorrect 
due to a mechanical failure
Mechanical damage to the sensor. 
Loss of fill fluid from the cell; 
ruptured diaphragm, broken 
sensor….
The Sensor must be replaced
Static Pressure Sensor 
Fail
The sensor signal value is incorrect 
due to a mechanical failure. Valid 
only for Differential pressure 
models
The circuitry for the sampling of the 
static pressure has failed.
The Sensor must be replaced
Device Troubleshooting
...Diagnostic, BlOCK_ERR and Status Byte