160 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Deleting a measurement range
NOTE
The deletion of measurement ranges is only possible with the Fidas24 ana-
lyzer module.
Menu path
MENU Configure Component-Specific Measurement Range
Select Component ...
Selection
All measurement ranges configured (at the factory) for a sample component
are displayed.
If a measurement range can be deleted from the configuration of the sample
component, then the softkey DELETE MEASURING RANGE is displayed.
NOTES
The deletion of measurement ranges can be locked when configuring the
gas analyzer by means of the test and calibration software TCT.
The active measurement range (in which measurements are currently taken)
and measurement ranges of the currently active calibration method cannot
be deleted.
Procedure
1 Press DELETE MEASURING RANGE.
2 Confirm safety interrogation by pressing DELETE MEASURING RANGE, if
necessary, enter password (level 1).
The entry "Free" is displayed in the list instead of the deleted measure-
ment range.