110 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Configuring the QAL3 monitoring
ATTENTION
The QAL3 data stored on the memory card is lost if the ECT software tools
("configurator") or the TCT are used to change the configuration of the gas
analyzer or if the memory card is installed in another gas analyzer. In these
cases, the QAL3 data must be exported (see page 114) beforehand.
NOTES
The QAL3 monitoring is deactivated ex works for all sample components, i.e.
no QAL3 data is stored.
If a password (see page 102) has been configured in the gas analyzer, this
password must also be input in the configuration of the QAL3 monitoring.
Call up the home page of the QAL3 monitoring
1 Ensure that the gas analyzer is connected to a PC via the Ethernet port.
2 Open the Web browser on the PC, enter the IP address of the gas ana-
lyzer (if necessary, ask the network administrator) and establish the
connection.
The QAL3 monitoring home page of the EasyLine EL3000 Series is
opened.
Configuring the sample component
1 Click "QAL3 Monitoring" on the QAL3 monitoring home page.
2 Click "Components" (sample components) on the "QAL3 Monitoring"
page.
3 Click the sample component to be configured on the "Components"
page.
4 Enter the following data and values in the "Edit Component" page:
"Active" (QAL3 Monitoring of the sample component activated),
"Installation" (installation location)
"Technician",
"Comment",
"sAMS Zero" (S
AMS
value for zero point)
"sAMS Span" (S
AMS
value for reference point)
5 Exit the page by clicking "Save" or "Cancel".
Configuring the printer output
1 Click "QAL3 Monitoring" on the QAL3 monitoring home page.
2 Click "Settings" in the "QAL3 Monitoring" page.
3 Check and, if required, change the number of lines on the first page
("Number of lines on first page" - recommendation for portrait format:
35) in the "Printing" window, as well as on the following pages ("Number
of lines on all other pages" - recommendation for portrait format: 50).
4 Exit the page by clicking "Save" or "Cancel".

Table of Contents

Other manuals for ABB EasyLine EL3060 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3060 Series and is the answer not in the manual?

ABB EasyLine EL3060 Series Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3060 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals