EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 119
Carrying out a calibration reset
Menu path
Maintenance Basic Settings Calibration Reset
When should a calibration reset be carried out?
A calibration reset should only be carried out if the gas analyzer can no long-
er be calibrated by normal means. A possible cause is e.g. that the gas ana-
lyzer has been calibrated with the wrong test gases.
What does the calibration reset do?
The calibration of the gas analyzer is reset to the ex-works calibration set-
tings by the calibration reset. Furthermore, the offset drift and the amplifi-
cation drift are electronically reset to the ex-works calibration settings.
Status signal
The status signal "Maintenance mode" (see page 130) is active during the
calibration reset.
NOTE
The calibration reset cannot be carried out while an automatic calibration is
in progress.