136 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
No. Status Message Reaction/Troubleshooting
500 iQ An internal calibration error has occurred. Repeat the calibration. Check the configura-
tion with TCT. Notify Service.
501 iQ The requested functional capability is not
available in the instrument.
Check the configuration with TCT.
502 iQ A calibration error has occurred in the in-
strument.
Calibration will be terminated. Notify Ser-
vice.
503 W S iQ The sensitivity is too low! The calibration
has been terminated.
Incorrect test gas! Check the test gas supply
and
infeed; repeat the calibration.
508 iQ Unknown calibration error. Check the con-
figuration.
Message during the automatic calibration.
Check the configuration with TCT.
511 iQ Autocalibration terminated For information
512 F a Autocalibration is in progress. For information
513 iQ An internal calibration error has occurred. Repeat the calibration. Check the configura-
tion with TCT. Notify Service.
517 F a Instrument is in operation.
518 iQ The calibration could not be carried out
because the measured value is unstable.
Check the test gas supply and infeed; repeat
the calibration.
519 iQ The calibration could not be carried out
because the preamplifier is overdriven.
Check the test gas supply and infeed; repeat
the calibration.
529 W S iQ The calibration was stopped because no raw
measured values can be recorded.
Check the test gas supply and infeed; repeat
the calibration.
538 W S iQ The zero-point calibration was stopped,
because the analyzer is contaminated!
Clean the sample cell. Notify Service.
539 W S iQ The zero-point calibration was stopped,
because the analyzer is contaminated!
Clean the sample cell. Notify Service.
801 A S a An external failure has occurred. Input signal at the correspondingly config-
ured digital input.
802 W a An external maintenance request has oc-
curred.
Input signal at the correspondingly config-
ured digital input.
803 F a A user-defined maintenance mode occurred
during: