76 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Test gases
Analyzer Test gas for zero-point calibration
and single-point calibration
Test gas for end-point calibration
Magnos28 Oxygen-free process gas Process gas with known O
2
concen-
tration
Magnos28 with suppressed
measuring range
Test gas with O
2
concentration near
the starting point of the measuring
range
Test gas with O
2
concentration near
the end point of the measuring range
Magnos28 with single-point
calibration
Test gas with O
2
concentration in an
existing measuring range or ambient
air. Same moisture content as pro-
cess gas.
Magnos28 with substitute
gas calibration
Oxygen-free process gas or substi-
tute gas (O
2
in N
2
)
Substitute gas, e.g. dried air
CAUTION
In order to avoid accumulations of explosive gas mixtures, do not use air as
a test gas for single-point calibration when measuring flammable gases!
Dew point
The dew point of the test gases must be approximately the same as the dew
point of the sample gas.

Table of Contents

Other manuals for ABB EasyLine EL3060 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB EasyLine EL3060 Series and is the answer not in the manual?

ABB EasyLine EL3060 Series Specifications

General IconGeneral
BrandABB
ModelEasyLine EL3060 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals