12 | NGC8200 AND PGC1000 USER MANUAL | 2101510MNAH
Helium Carrier Consumption rate: 12 ml/minute typical to 20 ml/minute maximum
Medium 800 to 1500 Btu per standard Cubic Foot (29.8 to 44.6
megajoules/meter
3
) with less than 100 PPM H2S
Analysis time Approx. 5 minutes; interval between cycles is adjustable
Calibration/validation
streams
Up to 2 dedicated (reduces sample stream for each dedicated
calibration streams). Must use dedicated stream(s) for Auto-Cal
feature.
Sample streams Up to 4 (with manual calibration streams)
Construction NEMA/Type 4X (IP56)
Aluminum Alloy with white polyester powder coating.
Explosion-Proof, see specification sheet for certifications.
Installation time Requires 2-3 hours for installation, minimum 8 hours run time for
repeatability.
Mounting Pipe run, free-standing pipe, shelf, and environmental enclosure
NGC8200 and PGC1000
dimensions
Width Height Depth Weight
US 9.5 inches 8.82 inches 15.64 inches 29 lbs.
Notes:
1. Usually experienced at startup. Use this for power supply sizing requirements (includes approx.
20% buffer and is calculated for maximum allowable power supply voltages).
2. At recommended AC power supply voltage. Highly temperature dependent, with feed-through
heater operating continuously. Usually occurs at only the coldest ambient operating temperature, i.e.,
0 °F (-18 °C).
2.1 Analyzer standard hardware features
The analyzer features a rugged, field-ready design. Hardware features include:
Enclosure
Cast aluminum housing with six exterior hubs
Powder coating
Weatherproof construction
Modular design (see Figure 2-1)
Digital controller assembly
Analytical module with compact design and single bolt replacement
Feed-through assembly with flame path arrestors
Termination panel
Electronics
32-bit digital controlling electronics (no analog control loops)
Low power operation
Dual digital carrier pressure regulation
Digital temperature control
Digital detector electronics
Low EMI/RFI design
Operates on Windows CE
Auto-start with diagnostics
Factory calibrated