2101510MNAH | NGC8200 AND PGC1000 USER MANUAL | 55
Figure 6-17: PCCU setup for local Ethernet communication
7. Click Close to close connection setup.
8. Click Entry on the PCCU32 menu bar to connect to the device.
9. Click the Entry icon on the PCCU32 toolbar to connect to the device. If unable to connect, see
section 8.6 Basic Ethernet connection troubleshooting.
Figure 6-18: PCCU32
10. Click Yes if the message to synchronize the date and time displays. This message displays if
the analyzer’s calendar clock does not match the laptop's date and time which is usually the
case with a new system.
When connection is successful, the analyzer’s startup wizard screen displays. The wizard provides the
sequence to complete basic analyzer configuration. Go to section 6.5 Startup wizard.
6.4 Power on sequence
6.4.1 Initialization
At power on, the analyzer starts its initialization sequence:
Performs a cold-start
Loads startup information into RAM memory. The message “Boot Loader” displays.
Start Diagnostics. Diagnostic results are available on the Diagnostic tab in PCCU. Once the
initializing sequence is complete, connect with the analyzer using PCCU.
6.4.2 Startup diagnostics
The analyzer has an extensive built-in list of tests which are automatically performed each time the
device is started.