2101510MNAH | NGC8200 AND PGC1000 USER MANUAL | 59
6.5.1 Set up the station
1. At the Startup Wizard, click Next. The Station Setup tab displays (Figure 6-23).
Figure 6-23: Station setup
2. Enter the Station ID (10 alphanumeric digits) and Location (24 alphanumeric digits). See
3. Table 6-4. The Station ID should be a unique identifier from other analyzers.
4. Verify the date and time; if incorrect, select Yes for Set Device with the PCCU Date/Time.
IMPORTANT NOTE: Additional items in the Station Setup screen are not required for startup. For
more information regarding the setup of these items, click Help to see more information about
each screen.
5. When all desired changes have been made, click Send and then Next to move to the next
screen.
Table 6-4: Station setup basic parameters
Description
Value
Station ID
Assign unique
i
dentifier (10 alphanumeric digits)
Location Enter meter location (24 alphanumeric digits).
Date/Time Shows current device Date/Time
Set device with PCCU Date/Time Resets device date and time to match PC/laptop
6.5.2 Set up the stream
1. At the Stream Setup tab (Figure 6-24), enter the Stream ID, Location, Contract Settings, and
Calculation Settings (see
2.
3. Table 6-5).