2101510MNAH | NGC8200 AND PGC1000 USER MANUAL | 69
Figure 6-36: Analyzer Operation screen
If completed satisfactorily, the startup Wizard should not re-appear when connecting to the analyzer
again. However, should you want to review or make changes, you may re-start the Wizard by
selecting the Wizard icon from the PCCU32 top menu bar (Figure 6-37).
Figure 6-37: Restart the Startup Wizard from the Operation screen
6.6 Manual peak find
Many troubleshooting procedures require the Peak Find tools. With Peak Find you can examine
chromatograms to determine if component peaks appear as expected, or modify certain parameters
when chromatograms do not display expected results.
IMPORTANT NOTE: This section describes procedures using peak find in Manual mode. This
mode allows user configuration of relevant parameters if fine tuning/correction is needed. Manual
mode is the only available mode and is automatically used in systems with versions of Flash
2102411-043 and OS 2102141-016 and later. Earlier software versions show options for either
Manual or Automatic modes. For systems with those software versions, manual mode is still
recommended and must be selected to complete the procedures in this section.
Peak Find consists of the following (Figure 6-38):
The Peak Setup and Peak Table tabs to determine or modify the peak characteristics and
label peaks when needed.
Pressure and time/duration parameters for sample processing
Function buttons to save, process peak setup changes, or run an analysis cycle when required.