• It is the operator’s responsibility to protect the patient against unnecessary or excessive radiation doses.
• Additional protection devices (aprons, collars, etc) are required to protect the patient from radiation.
• Before exposing patients with pacemakers, contact the manufacturer of the latter to ensure that the X-rays generated by X-MIND unity
do not interfere with its functionality
• X-MIND unity generates x-rays: Before using this x-ray system please refer to the regulation in force in your area concerning paediatric
patients, pregnant women and anyone with health issues that contraindicate the use of x-rays. Investigate and make sure of this condi-
tion before starting the exposure.
This symbol indicates X-ray hazards
• According to the paragraph 203.8.5.4 of the IEC 60601-2-65, in case you’re using the X-MIND unity device together with the integrated
Sopix/Sopix2 Inside sensor, it is mandatory to install and use the original rectangular beam limiting device provided by the manufactu-
rer of the X-MIND unity.
CAUTION
MECHANICAL RISK
• Pay extreme attention to the tension of the internal spring of the scissor arm in order to avoid the arm opening outward and causing
injury.
• Check that the installation of the unit complies with the mechanical specications of the support (walls, ceiling, etc..) where it is instal-
led
• Adjustments or any kind of attempt of repairing or disassembling must only be performed by qualied and authorized service personnel.
• The X-MIND unity must not be used in environments or close to environments subjected to mechanical vibrations or mechanical shocks.
• Do not cause overbalancing of the Mobile Version due to improper pushing or leaning the device using other parts respect to the handle,
or with the brakes activated, or with the scissor arm opened.