EasyManua.ls Logo

Acteon X-MIND unity - Check the Selected X-Ray Tube Voltage; Check the Selected X-Ray Tube Current; Check the Selected Type of Patient

Acteon X-MIND unity
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 31 of 79
ENGLISH
Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
5.3. CHECK THE SELECTED X-RAY TUBE VOLTAGE
The set x-ray voltage will be shown on the display
To change the selected value press “kV” key until you have reached the desired value
It is possible select x-ray voltage among these values: 60kV/65kV/70kV
5.4.CHECK THE SELECTED X-RAY TUBE CURRENT
The set x-ray current will be shown on the display
To change the selected value press “mA” key until you reached the desired value
It is possible select x-ray current among these values: 4-7mA (±1mA step)
4 mA
5 mA
6 mA
7 mA
5.5.CHECK THE SELECTED TYPE OF PATIENT
The led (at the bottom of the key) relevant to the selection of the desired patient size should be illuminated or
led CHILD ON indicates that the x-ray system is set for a patient with a small physique
led ADULT ON indicates that the x-ray system is set for a patient with a large physique
Alternately pressing this button switches from child to adult settings.
PLEASE NOTE
After the modication, the default exposure values will be automatically changed

Table of Contents

Other manuals for Acteon X-MIND unity

Related product manuals