X-RAY CALIBRATION PROCEDURE
CAUTION
Perform this operation only if necessary or suggested by the Technical Support Service
During this operation x-rays will be emitted!
It is mandatory to adopt all the safety measures relevant to radioprotection.
1. To perform an X-ray tube calibration procedure please see the instructions below: Press and hold key for 5 seconds until the
message “TUBE CALIBRATION” appears on the display
2. Keeping a safe distance away from the x-ray beam press and hold in. the exposure key .
The “CAL. IN PROGRESS” message will appears on the display. Exp. Led is illuminated and an intermittent acoustic sound is emit-
ted.
3. At the end of the procedure, the acoustic sound stops and the “CAL. SUCCESS” message appears on the display.
4. When the pause light goes out the control unit will resume back to standard functioning mode and is ready for a new exam
PLEASE NOTE
The calibration procedure takes about 50s
After this procedure, the PAUSE led will start blinking. During this period all the functions of the equipment are inhibited. Please
wait until the X-MIND unity completes the cooling time due to the tube calibration before resuming normal operation.