Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
ENGLISH
CUSTOMIZE DEFAULT EXPOSUR
VALUES
PLEASE NOTE
The value and steps of the 21 programmed exposure times of the X-MIND unity (refer to chapter 3 ) cannot be modied in the X-MIND unity
x-ray system, since their values are in conformity according to the regulation in force concerning x-ray intra-oral equipment. however the
operator does have the abilityto customize the default exposure values.
!
WARNING
The customizing of the exposure values is just applied to lm or sensor type which is currently selected.
After customizing, the “Charts of default exposure values” (refer to Chapter 5) for that specic lm or sensor type are NOT valid anymore;
however, for all the other lm or sensor types for which no customizing has been applied, the default exposure values in the charts are still
valid.
To programm the new exposure values of the lm or sensor type that iscurrently selected, press either of the following keys:
Thebelow symbols being illuminated shown on the display represent the possibility to modify the exp. values scale:
PLEASE NOTE
When either the keys are pressed and held in the “repeat” function will cause the displayed time to scroll faster.
For conrming the new program check the led status of the key
led BLINKING MEMO indicates that it is possible to save the new default exposure value. Press the button to save the new
default exposure value.
The MEMO OFF led indicates that is not possible to save the new default exposure value.
PLEASE NOTE
It is not possible to save data when the “range of exposure eld” exceeds the programmed exposure time limits.
77