ENGLISH
Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
5.13.1. SUGGESTED INCLINATION ANGLES
MAXILLA
Molar 35°
Premolar and canine 45°
Incisor 55°
Bite-wing exam 10°
-----------------------------------------------------------
Bite-wing exam
Incisor -20°
Premolar and canine -10°
Molar -5°
MANDIBLE
5.14. MAKE THE EXPOSURE
PLEASE NOTE
To obtain the best radiological diagnostic results, pay extreme attention in all the steps of the process: positioning the patient, the x-ray
generator and collimator, exposing the x-ray detector (lm/digital sensor/plates); and processing the lm or the digital image acquisition
and processing chain.
CAUTION
It is mandatory for the operator to check the parameters before making any exposure with either the local or remote exposure switch.
CAUTION
Always maintain audio and visual contact with the patient and unit during the x-ray exposure: the operator is required to monitor the
patient and the X-Mind unity parameters throughout the entire duration of the x-ray examination.