Operator’s manual • X-Mind unity • VH • (15) • 09/2017 • NUN0EN010H
The following exposure times (s) have been stored:
0,020 – 0,025 – 0,032 – 0,040 – 0,050 – 0,063 – 0,080 – 0,100 –0,125 – 0,160 – 0,200 – 0,250 – 0,320 – 0,400 – 0,500 –0,630 – 0,800 – 1,00 –
1,250 – 1,600 – 2,000 s
PLEASE NOTE
These times are in compliance with IEC 60601-2-7 standard according to the 2nd and 3rd edition of the IEC 60601-1 and with the ISO 497 series
R’10 recommendations.
PLEASE NOTE
These values of the programmed exposure times MAY NOT be modied.
Certain exposure values have been predened which depend on the selection of the operating parameters:
- cone (8”/12”)
- type of patient (ADULT/CHILD)
- x-ray technique
- intra-oral exam type
PLEASE NOTE
If one so desires, it is possible to change these values by means of the dedicated key buttons previously indicated.
Possible modications of the exposure values:
- x-ray anodic voltage (60kV/65kV/70kV)
- x-ray anodic current (typical: 4mA/6mA/7mA)
- type of patient (ADULT/CHILD)
- x-ray technique
—> refer to chapter 5
Possible modications of the parameters:
- Number of exposure switches
- Cone size (20cm - 8” / 30cm - 12”)
- Cone type (Square / Round)