8230 Optical Power Meter Operation Manual
8.1.2 General Specifications
8-3
8.1.2 General Specifications
8.1.3 Standard Accessories
8.1.4 Option
Item Specification
Operating environment Environmental temperature: 0 to +40C.
Relative humidity: 80% or less with no condensation
Storing environment Environmental temperature: -20 to +70C.
Relative humidity: 80% or less with no condensation
Warm-up time 30 minutes or more (to achieve prescribed accuracy)
Power supply AC adapter: AC100 V-240 V, 0.29 A-0.16 A, 50/60 Hz
Power supply frequency 50/60 Hz
Power consumption 100 to 120 V: 5 VA or less, 220 to 240 V: 10 VA or less
AC adaptor rating 100-240 V, 0.29 A-0.16 A, 50/60Hz
DC input 9 V 100 mA or less
Battery drive 4 AA size batteries *1
Battery life: 60 h (with 1 mW or less power, with back light OFF, using
the alkaline dry battery, and at +23 C 5 C)
External dimensions Approx. 80 (width) x 180 (height) x 40 (depth) mm
Mass 300 g or less (excluding AA batteries)
*1: Batteries are not included.
Item Model Quantity Remarks
AC adaptor A146001 1 AC100V-240V
Operation manual E8230 1 English
Option to use old-type sensors OPT8230+70 821x series:
The 82014A, 82017A, (82015,or Q82018A *1 )
can be used.