EasyManuals Logo

ADCMT 8230 User Manual

ADCMT 8230
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
8230 Optical Power Meter Operation Manual
7.2 Precautions
7-2
7.2 Precautions
1. When calibrating, avoid any dust, impact, and noise, and perform the calibration under the following
environmental conditions:
Temperature 23 C 5 C, temperature change of 1 C
Relative humidity 70% or less
2. This instrument requires a warm up of 30 minutes or more. Sufficient warm-up time is also required for
the measuring instrument before starting calibration.
The 6245 requires a warm up of one-hour or more.
3. Calibration cannot be performed by using the key pad. Perform calibration by remote control through
the USB.
4. After the calibration is complete, it is recommended to write the dates of the calibration and the next
calibration on the sticker and affix it to the instrument.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADCMT 8230 and is the answer not in the manual?

ADCMT 8230 Specifications

General IconGeneral
BrandADCMT
Model8230
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals