8230 Optical Power Meter Operation Manual
4.6.2 Remote Control Command List
4-21
Instru-
ment cali-
bration
Offset data direct 
setting (on RAM)
XOF rng, 
data
rng: Range Setting range: 4 to 11
data: Off Set Data
Setting range : 20000
*6
Reading the 
offset data 
(on RAM)
XOF? rng rng: Range Setting range: 4 to 11
Reply:XOFr, ddddd
*6
Saving the cali-
bration data
XWR Writes the calibration data into the non-volatile 
memory.
*6
Calibration data 
initialization
XINI rng rng: Range Setting range: 0, 4 to 11
if setting to 0, all ranges are initialized.
IV offset calibration data is not initialized.
*6
*6: Available only when the calibration mode is set.
Table 4-5    Remote Control Commands (7 of 7)
Item Command Description
Initial value Availability
Power 
ON
Factory 
default
Calibration 
mode
Error2