8230E Optical Power Meter Operation Manual
1.4 Connectable Sensors
1-3
1.4 Connectable Sensors
Thin- and cylindrical-type 823x series sensors can be connected to this instrument.
Functions of the sensors are categorized into general purpose and high power.
82311B
The 82311B thin-type, general-purpose optical sensor is available for measurement in optical disk
equipment or narrow spaces. Wavelengths ranging from 390 nm to 1100 nm can be measured and the
standard calibration wavelength is 780 nm. The calibration wavelength addition option (405 nm:
OPT82311B+21 or 650 nm: OPT82311B+22) can be added.
For the standard model, the typical value for wavelength sensitivity correction is saved in the memory
of the sensor. The actual measurement value for the sensor-specific wavelength sensitivity correction
can be acquired by specifying the wavelength sensitivity correction option (OPT82311B+20).
82321B
The 82321B cylindrical-type optical sensor is available for attachment to test equipment, for measure-
ment on an optical bench, or for optical fiber system measurement. Adapters to accommodate each op-
tical connector are available. Wavelengths ranging from 390 nm to 1100 nm can be measured and the
standard calibration wavelength is 780 nm. The calibration wavelength addition option (405 nm:
OPT82321B+21 or 650 nm: OPT82321B+22) can be added.
For the standard model, the typical value for wavelength sensitivity correction is saved in the memory
of the sensor. The actual measurement value for the sensor-specific wavelength sensitivity correction
can be acquired by specifying the wavelength sensitivity correction option (OPT82321B+20).
82313B
The 82313B thin-type optical sensor is available for measurement in optical disk equipment or narrow
spaces. The 82313B is ideal for measuring optical disk equipment in which a high power laser diode
(LD) is used. If the light condenses on the incident surface, up to 200 mW of optical power can be mea-
sured without saturation. Wavelengths ranging from 390 nm to 1100 nm can be measured and the stan-
dard calibration wavelength is 650 nm.
The calibration wavelength addition option (405 nm: OPT82313B+21 or 780 nm: OPT82313B+23) can
be added. The actual measurement value for wavelength sensitivity correction is saved in the memory
of each sensor.
82323B
The 82323B cylindrical-type optical sensor is available for attachment to test equipment, for measure-
ment on an optical bench, or for optical fiber system measurement. Adapters to accommodate each op-
tical connector are available. The 82323B is ideal for measuring optical disk equipment in which a high
power laser diode (LD) is used. If the light condenses on the incident surface, up to 200 mW of optical
power can be measured without saturation. Wavelengths ranging from 390 nm to 1100 nm can be mea-
sured and the calibration wavelength is 650 nm. The calibration wavelength addition option (405 nm:
OPT82323B+21 or 780 nm: OPT82323B+23) can be added. The actual measurement value for wave-
length sensitivity correction is saved in the memory of each sensor.
When option 70 is specified, the following sensors for the TQ8210 and TQ8215 can be used:
82014A
The 82014A is a general-purpose, cylindrical-type sensor used for the TQ8210 and TQ8215. When this
sensor is used for the 8230, option 70 must be installed. Wavelengths ranging from 400 nm to 1100 nm
can be measured and the calibration wavelength is 850 nm. The typical value stored in the optical power
meter is used as wavelength sensitivity correction data.