8230E Optical Power Meter Operation Manual
8. SPECIFICATIONS
8-1
8. SPECIFICATIONS
8.1 8230E Specifications
All accuracies are guaranteed for one year at a temperature of +23 C 5 C and a relative humidity of 70%
or less.
8.1.1 Instrument Specifications
Item Specification
Resolution 1 pW (W is displayed), 0.01 dB (dBm is displayed)
Accuracy The following is added to the accuracy of each sensor. (Within 24 hours
after the offset zero is executed, unit: W, display: 4-1/2 digits)
823x series sensor 821x series sensor
20 nW range (2.55% +200 digits) (0.55% +200 digits)
200 nW range (2.15% +20 digits) (0.15% +20 digits)
2 W to 200 mW range (2% +7 digits) (0% +7 digits)
Display LCD with three-level backlight
Wavelength display: 4 digits
Power display: 4 1/2 digits (Unit; mW, W, nW, dBm, and dBr)
Bar graph display
Range switching 8 ranges: Automatic, manual, and remote
Measurement speed Approx. 5 times/second
Compatible optical sensor 823x series: 82311B/82313B/82321B/82323B/82311/82313/82321/
82323
821x series: 82014A/82017A (82015/Q82018A *1)
Wavelength sensitivity correction Automatically corrects the sensor wavelength sensitivity by setting
wavelength (in 1 nm steps)
For the 823x series optical sensor, each value stored in the sensor is used
as the wavelength sensitivity correction value.
(Sensor-specific data for the 82311/82311B option 20 and 82321/
82321B option 20, or the typical values for the 82311/82311B and
82321/82321B standard models are stored.)
For the 821x series optical sensor, the typical value stored in the optical
power meter is used. *2
Wavelength preset function Four wavelengths where wavelength sensitivity correction is performed
can be registered. (This function can be used only from the operation
panel.)
Smoothing function Moving average from 2 to 100 times
Offset zero Stores the sensor offset in memory and automatically corrects it.