At “Select Display mH/mC”, press START to display “[mH]<
Display *[mC]”.
At “[mH]< Display *[mC]”, the current setting is indicated by
the symbol, “*”. Press < or > to toggle the display readout between
m°H and m°C.
Press ENTER to store the new setting (or STOP to restore the origi-
nal setting).
14. Serial Number
At “Serial Number”, press START to display the instrument seri-
al number (“Serial#: 123”).
Press ENTER to return to the setup menu.
15. Product/Test
Product/Test is for factory use only and has no field function.
16. Reservoir Status
This menu item provides information when filling the heat transfer
fluid reservoir.
At “Reservoir Status”, press the START button. The display
will report the present state of the reservoir.
If the reservoir is empty, the display will be “Fluid Level =
EMPTY”. As the reservoir is filled, the display will change sequen-
tially to “Fluid Level = LOW”, “HALF” and “FULL”.
Press STOP to return to the setup menu.
29
Installation & Setup