OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
36
Figure 4.10: Cancel Calibration Screen
4.6.3 Successful calibration
Once calibration completes, the instrument displays a message about calibration success
(Figure 4.11), or failure. If successful, the instrument is calibrated and is ready to perform
accurate sample tests.
Figure 4.11: Calibration Successful Screen
If the calibration failed, the reason for the failure is displayed. For example, if calibration failed
and the message states, “Wrong Calibrator 1” (Figure 4.12), this indicates the calibration test
was set up incorrectly. If a failure occurs during the calibration test process the system will
display the error, and automatically terminate the test run. Follow the instructions and attempt
another calibration. If the error persists, refer to the troubleshooting section.
Figure 4.12: Calibration Error