OsmoPRO® Multi-Sample Micro-Osmometer User’s Guide
55
7.7.4 Levey-Jennings chart
To display a Levey-Jennings for a control, press CHART. (Figure 7.21)
The chart shows the data from the selected control. This data is based on the results from
sample tests of the selected control, when Sample ID is enabled. The results can be filtered by
standard deviation (SD) bands.
Bands. Select 1, 2, or 3 from the box labeled SD Bands in the lower left corner to filter
the data based on the selection.
Press EXIT to return to the Controls screen.
Figure 7.21: Levey-Jennings Chart
7.7.5 Setting control actions
This function is only available on the Controls screen when the user is logged in with a
supervisor access level.
To change the action settings for the controls, press ACTION (Figure 7.22). The supervisor can
select what should occur if a control test result is outside of the alarm limits.
NOTE: Control action settings will only function if Sample ID is required for testing.
Figure 7.22: Action