AUTOMATIC
20
5.2.2 Recharge procedure
A Place the charger on a perfectly horizontal shelf and firmly attached to the floor
B Connect the power supply (SAV-C0013) to the charger and then to the power outlet
C The LED on the charger will flash green, indicating that it is ready to charge
D Insert the battery to charge into the battery charger as shown in image (14)
The recharging station allows you to recharge exclusively original ACC rechargeable Li ion batteries (SAV-C0011) of
A.M.I. Italia S.r.l. The charging time of around 2.5 hours may increase in the case of batteries that have undergone
recharging cycles higher than the one indicated. The battery charger is equipped with a control LED that indicates both
its functional status and the battery charge level, if inserted. The following is a table that allows identification of the
control LED coding:
Battery charger waiting for
battery insertion
When recharging, the battery charger control LED will flash green with a different frequency depending on the level of
recharge, until the charge is fully indicated by the control LED with FIXED green light.
Number of consecutive flashes
Defibrillation PADs
The Saver One allows the use of two different defibrillation PADs depending on the patient to be treated:
Defibrillation PADs for Adults model SAV-C0846
Defibrillation PADs for Children model SAV-C0016