AUTOMATIC
43
11.9 Patient analysis system
Determines the patient's impedance and evaluates the ECG rhythm and signal quality to
determine whether the shock delivery is appropriate.
≥4 seconds (with new fully charged battery) respecting IEC/EN 60601-2-4
97% Respects the guidelines IEC/EN 60601-2-4 2002(3) sources AHADB, MITDB and EDB
99% Respects the guidelines IEC/EN 60601-2-4 2002(3) sources AHADB, MITDB and EDB
If used on a patient who has the characteristics listed in the usage criteria, the Saver One
defibrillator is designed to recommend a defibrillating shock when it detects the right
impedance and when the following situations occur:
Ventricular Fibrillation peak-to-peak amplitude at least 200µVolts
Ventricular Tachycardia with cardiac rhythm frequency min. 180 bpm and peak-to-peak
amplitude at least 200µVolts (including ventricular flutter and polymorphic ventricular
tachycardia)
The Saver One is designed to not recommend shocks with all other rhythms, including: normal
sinusoidal rhythm, moderate ventricular fibrillation (<200 μVolts), some slow ventricular
tachycardia and asystoles.
11.10 ECG Analysis Function
Shockable rhythm
Ventricular Fibrillation (VF)
Shockable rhythm
Ventricular Tachycardia (VT, bpm>140)
Non-shockable rhythm
Normal sinusal rhythm
Non-shockable rhythm
Asystole
Untreatable rhythm
generic AF, SVT, PVC
Positive predictive values
11.11 Non rechargeable battery
Li-SOCl
2
(lithium-thionyl chloride) disposable, non-rechargeable
300 cycles of complete rescues (shocks at 200J. and CPR) *
200 cycles of complete rescues (shocks at 360J. and CPR) *
ECG analysis 35 continuous hours*
Duration in Standby
(installed battery)
4 years if installed in the AED, assuming an activation test, daily self-tests without turning on
the AED*
* New and fully charged battery at a constant temperature of 20°C and relative humidity without condensation 45%