AUTOMATIC
44
11.12 Rechargeable battery
Category
Nominal specifications
REF (Model)
SAV-C0011
Type
Li ion (lithium ions) Rechargeable
Voltage
21,6 VDC - 2100 mAh
Capacity
Standard 200J
250 continuous shocks*
Power 360J
160 continuous shocks *
ECG analysis
21 continuous hours*
Charging time
2,5 hours with charging station SAV-C0012*
Shelf Life
2 years or 300 charge/shock cycles (the one that occurs first) *
* New and fully charged battery at a constant temperature of 20°C and relative humidity without condensation 45%
11.13 Internal back-up battery
Category
Nominal specifications
Type
Battery Coin Cell (LiMnO2)
Purpose
Maintaining configuration data (date / time, etc.)
Voltage
3 VDC
Duration
Maintains data for 3 years (without external battery)
Maintains data for 6 years (with external battery inserted within 12 months)
11.14 Rechargeable battery charger
Category
Nominal specifications
REF (Model)
SAV-C0012
Charge control
LED multicolour red green (see paragraph 4.2.1)
Power supply
Input
15Vdc-2.67A / 12Vdc-5.5A
Output
26VDC 1,5A
Absorption
40W/66W
AC/DC Adapter
Model
MeanWell GS40A15-P1J
Identification code
SAV-C0013
Input
100-240VAC 50/60Hz 1.5A
Output
15V 2.67A
Absorption
40W