6 Auto test
The Saver One has been designed to be a totally safe device, always ready for use and able to automatically and
constantly verify correct operation, minimizing maintenance operations.
The Saver One performs different types of self-tests:
Activation: Every time a battery is inserted in the device
Automatically: During Stand-by mode with daily/monthly/half -yearly intervals
Switching on: When the device is switched on
The outcome of the control test can be viewed via a two-colour LED (green/red) and the LCD mini-display.
The mini-display and the control LED let you know at any time, even when the device is switched off (stand-by mode),
the functional status of the device and its battery.
6.1 Display and control LED
Both the mini display and the control LED are positioned on the front of the Saver One keyboard.
Based on the different colour of the control LED and the information shown on the display, the operator can
independently determine the functional status of the defibrillator and its battery.
The following table shows the flashing code of the control LED and the relative screens of the control mini display.