10.5 Troubleshooting Guide
The following table lists the symptoms, the possible causes and the possible corrective actions of the problems that may
arise. For more information on the implementation of corrective actions, refer to the other sections of the user's manual.
If the failure of the unit persists, request assistance.
Device with battery installed does
not switch on, the LED and the
control display are both off.
The battery is totally discharged
or faulty
Try replacing the battery. If the
problem persists, call for assistance
In standby the control LED
flashes green but the mini display
is off
The mini display is broken
Contact the assistance centre
In standby the control LED is off
but a "V" appears on the control
mini-display.
The control LED is broken
Contact the assistance centre
In standby the control LED
flashes RED and a wrench
appears on the control display.
During the daily self-test a critical
error of the device was found
Contact a service centre and report
the error code.
In standby the control LED
flashes GREEN / RED alternately
and a wrench appears on the
control display.
Very low battery
Level <1%
The device may turn off during
use.
(see the relevant paragraph)
In the operating mode the voice
command "Low battery" is
issued.
Low battery.
5% battery level.
It is possible to use the device but
the battery level is low
(see the relevant paragraph)
Get a new battery and replace it as
soon as possible.
During normal use the voice
command "Battery low, Replace"
The battery is depleted.
Level <1%
The device may turn off during
use.
(see the relevant paragraph)
Avoid using the device if possible.
Replace the battery
With the device turned on and
after placing the PADs on the
patient, the device continues to
communicate:
"Place electrodes"
The PADs connector has not been
inserted correctly or it has been
removed
Insert the PADs connector in the
appropriate compartment
The PADs have been placed
incorrectly
Correctly position the PADs on the
patient's stripped chest. If
necessary, remove the hair from the
chest with a razor
PADs are not working properly
Check the integrity and expiration
of the PADs, replace them if
necessary
The device turns on, the mini is
on but no voice command is
issued
The device's speaker does not
work