AUTOMATIC
35
10 Maintenance
The Saver One defibrillator was designed to make maintenance operations as simple and autonomous as possible. In
fact, thanks to the control tests carried out in total autonomy by the device, it is not necessary to perform any
extraordinary maintenance, but only routine maintenance, which consists of a frequent visual check of the LED and the
control display, together with a visual inspection of the relative accessories.
Whenever it is necessary to contact the supplier for assistance during an installation, or to report anomalies, contact the
supplier using the references:
Assistance request email: info@amiitalia.com
Phone: +39 081 806 05 74
Website: www.amiitalia.com
10.1 After each use
After using the Saver One defibrillator it is necessary to proceed with the following operations in order to prepare the
device for the next use:
1. Check the presence of the memory card and its remaining capacity (see paragraph 5.3)
2. Check that the control LED is on with flashing lighting (flashing green)
3. If they have been used, replace the PADs with a new package
4. If not used, check the expiry date of the PADs, if expired replace them with a new package
10.2 Ordinary maintenance
Thanks to the control tests carried out in total autonomy by Saver One, ordinary maintenance will require a simple and
quick inspection, following the operations described in the table:
Check
Daily
Check
Monthly
Check
before use
Check
after use
Action indicated
*
*
*
Check the LED and the control display.
*
*
*
Check the integrity of the device, its parts and the accessories supplied.
*
*
Check the expiration date of the defibrillation PADs.
*
*
Check the remaining capacity of the memory card.