AUTOMATIC
31
8.5 Shockable rhythm
If a Ventricular Fibrillation or Tachycardia is detected, the Saver One will inform the operator using the following
commands:
Voice commands
Display
Icons/ Luminous buttons
Shock recommended
Do not touch the
patient icon on fixed
light
Keep distance, charging
Auto Shock icon
flashing in red
Keep distance, the shock will be
automatically delivered in 5 (five)
seconds
5 seconds countdown starts
At the end of the countdown the Saver One will perform the defibrillation shock.
In this phase the AUTO shock indicator will no longer flash and the device will inform the operator using the following
commands:
Voice commands
Display
Shock delivered
The Saver One delivers the shock using the BTE waveform with auto compensation of the patient's thoracic impedance.
The Saver One download protocol is incremental, i.e. the energy delivered to the patient varies incrementally based on
the number of shocks performed:
Saver One 200J: The first shock is delivered to energy of 150J the following at 200J
Saver One 360J: The first shock is delivered to energy of 200J the second at 250J the following at 360J
The detected impedance value must be between 20 and 200 Ohm; if a value outside this range is detected, it is required
to position the PADs.
The shock protocol is pre-set, cannot be modified by the user and is reset at each power up.
AUTOMATIC SHOCK
IN FIVE SECONDS