Network protocols (TCP/IP services) FTP, Telnet, HTTP, SNMP, LPD
Image formats DICOM (Default)
TIFF
Power consumption - heat dissipation
During operation 450 W - 1620 kJ/h
In standby 200 W - 720 kJ/h
Peak power (absolute max. rating) 700 W - 2520 kJ/h
Protection against
Electrical shocks Class 1 (grounded)
Ingress of water IPXØ
Environmental conditions (operation)
Room temperature Between +15°C and +30°C
Relative humidity Between 20% and 75%
Note: Films may not become wet!
Atmospheric pressure 70 kPa - 106 kPa
Related altitude on site 3000 m to 0 m
Environmental storage conditions
Note: Climate conditions for storage are in accordance with
EN60721-3-1-class 1K4.
Room temperature Between -25°C and 55°C (storage)
Relative humidity Between 10% and 100%
Absolute humidity Between 0.1 g/m
3
and 35 g/m
3
Rate of change of temperature 1°C/min
Atmospheric pressure 70 kPa - 106 kPa
Environmental transport conditions
14 | Drystar 5500, Drystar 5503 | Drystar 5500, Drystar 5503
2901J EN 20191018 0826