Agilent 1260 Infinity DAD and MWD User Manual 49
Installing the Module
3
Installing the Detector
1 Note the MAC address of the LAN interface (rear of the module, under the
configuration switch, see Figure 9 on page 49). It’s required for “LAN
Configuration” on page 261.
2 Place the module in the stack or on the bench in a horizontal position.
3 Ensure the line power switch at the front of the module is OFF.
4 Connect the power cable to the power connector at the rear of the module.
Figure 9 Rear View of Detector
5 Connect the CAN cable to other Agilent 1200 Series modules.
6 Connect the LAN cable (e.g. from a Agilent ChemStation as controller) to
the detector's LAN connector.
7 Connect the analog cable(s) (optional).
8 Connect the APG remote cable (optional) for non-Agilent 1200 Series
instruments.
3RZHU
NOTE
In multi-detector configurations the LAN of the G1315C/D and G1365C/D must be used
due to its higher data load.