Agilent InfinityLab LC Series RID User Manual 116
8 Test Functions
Optical Balance
4
Adjust
Optical Balance
.
a
While monitoring the optical balance use the flat-headed screwdriver to
turn the zero glass adjustment screw slowly.
b
When the diode balance value reaches 0.00 (green area), optical balance is
restored.
Figure 21 Turning the zero glass adjustment screw
Zero glass adjustment
(screw left)