Agilent InfinityLab LC Series RID User Manual 96
7Error Information
Refractive Index Detector Specific Error Messages
Undecipherable Temperature Signal
Error ID: 2684
Maximum Temperature Exceeded
Error ID: 2685
The maximum heater temperature has been exceeded.
Probable cause Suggested actions
1
Heater cable disconnected.
Please contact your Agilent service representa-
tive.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective heater.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective heater.
Please contact your Agilent service representa-
tive.